本文主要研究内容
作者胡志纲,任建,张红(2019)在《医用电子直线加速器的性能测量及不同指形电离室对其剂量率的响应特性分析》一文中研究指出:目的:分析医用电子直线加速器的性能测量及不同指形电离室对其剂量率的响应特性分析。方法:选取宝鸡市5家医疗机构正常运行的9台医用加速器(3台国产、6台进口),依照国家标准和职业卫生标准,检测医用电子直线加速器的性能,进行剂量精度及机械性能抽样检测;并选择IBA 0.01 cm3、PTW 0.125 cm3和0.6 cm3的3种不同指形电离室,以100~600 cGy/min剂量范围加速器,在同等机器跳数(MU)与测量条件下,分析不同指形电离室剂量率的响应特性。结果:检测的9台医用加速器检测率为100%;电子线检测中9台医用加速器的穿透特性范畴为-6.54%~5.75%、吸收剂量指示值相对偏差为-9.96%~-2.80%,其合格率分别为66.67%和44.44%,相对较低。X射线检查中,国产医用直线加速器1台灯光野与照射野偏差为3.1 mm,1台吸收剂量指示值相对偏差为4.05%;进口设备1台穿透特性为-9.11%,经调试后3台仪器检测合格率均为100%。在相同能量前提下3种指形电离室中剂量率响应下降幅度最大为IBA 0.01 cm3电离室,6 MV时下降1.60%,15 MV时下降1.40%;PTW 0.125 cm3电离室6 MV时下降0.50%,15 MV时下降0.40%;PTW 0.6 cm3电离室6 MV时下降0.70%,15 MV时下降0.30%。结论:对医用电子直线加速器进行性能测量有助于及时修正剂量偏差;医用电子直线加速器的剂量采集效率会受到剂量仪、不同指形电离室的剂量率与工作电压影响,在剂量测量时应分析、测试设备剂量率,提高仪器精度、检测率及合格率。
Abstract
mu de :fen xi yi yong dian zi zhi xian jia su qi de xing neng ce liang ji bu tong zhi xing dian li shi dui ji ji liang lv de xiang ying te xing fen xi 。fang fa :shua qu bao ji shi 5jia yi liao ji gou zheng chang yun hang de 9tai yi yong jia su qi (3tai guo chan 、6tai jin kou ),yi zhao guo jia biao zhun he zhi ye wei sheng biao zhun ,jian ce yi yong dian zi zhi xian jia su qi de xing neng ,jin hang ji liang jing du ji ji xie xing neng chou yang jian ce ;bing shua ze IBA 0.01 cm3、PTW 0.125 cm3he 0.6 cm3de 3chong bu tong zhi xing dian li shi ,yi 100~600 cGy/minji liang fan wei jia su qi ,zai tong deng ji qi tiao shu (MU)yu ce liang tiao jian xia ,fen xi bu tong zhi xing dian li shi ji liang lv de xiang ying te xing 。jie guo :jian ce de 9tai yi yong jia su qi jian ce lv wei 100%;dian zi xian jian ce zhong 9tai yi yong jia su qi de chuan tou te xing fan chou wei -6.54%~5.75%、xi shou ji liang zhi shi zhi xiang dui pian cha wei -9.96%~-2.80%,ji ge ge lv fen bie wei 66.67%he 44.44%,xiang dui jiao di 。Xshe xian jian cha zhong ,guo chan yi yong zhi xian jia su qi 1tai deng guang ye yu zhao she ye pian cha wei 3.1 mm,1tai xi shou ji liang zhi shi zhi xiang dui pian cha wei 4.05%;jin kou she bei 1tai chuan tou te xing wei -9.11%,jing diao shi hou 3tai yi qi jian ce ge ge lv jun wei 100%。zai xiang tong neng liang qian di xia 3chong zhi xing dian li shi zhong ji liang lv xiang ying xia jiang fu du zui da wei IBA 0.01 cm3dian li shi ,6 MVshi xia jiang 1.60%,15 MVshi xia jiang 1.40%;PTW 0.125 cm3dian li shi 6 MVshi xia jiang 0.50%,15 MVshi xia jiang 0.40%;PTW 0.6 cm3dian li shi 6 MVshi xia jiang 0.70%,15 MVshi xia jiang 0.30%。jie lun :dui yi yong dian zi zhi xian jia su qi jin hang xing neng ce liang you zhu yu ji shi xiu zheng ji liang pian cha ;yi yong dian zi zhi xian jia su qi de ji liang cai ji xiao lv hui shou dao ji liang yi 、bu tong zhi xing dian li shi de ji liang lv yu gong zuo dian ya ying xiang ,zai ji liang ce liang shi ying fen xi 、ce shi she bei ji liang lv ,di gao yi qi jing du 、jian ce lv ji ge ge lv 。
论文参考文献
论文详细介绍
论文作者分别是来自中国医学装备的胡志纲,任建,张红,发表于刊物中国医学装备2019年04期论文,是一篇关于医用电子直线加速器论文,指形电离室论文,性能测量论文,剂量率论文,中国医学装备2019年04期论文的文章。本文可供学术参考使用,各位学者可以免费参考阅读下载,文章观点不代表本站观点,资料来自中国医学装备2019年04期论文网站,若本站收录的文献无意侵犯了您的著作版权,请联系我们删除。
标签:医用电子直线加速器论文; 指形电离室论文; 性能测量论文; 剂量率论文; 中国医学装备2019年04期论文;