作者陈彦(2019)在《迈瑞PM9000多参监护仪的NIBP模块维修三例》一文中研究指出:无创血压是利用非直接方式测量血压,最常用的两种方法分别是柯氏音法和脉搏震荡法[1]。迈瑞PM9000监护仪的自动化无创性测压(automated non-invasive blood pressure,NIBP)模块使用的是脉搏震荡法,其原理是通过气泵给袖带充气,直到袖带压力阻断上臂动脉血流,然后按照一定的幅度放气,阻断的动脉血流随着压力的减少和脉搏跳动,在袖带中产生脉动,袖带连接到的压力传感器记录其信号并放大,再转换为数字信号,通过单片机记录和算法处理得出收缩压和舒张压
mo chuang xie ya shi li yong fei zhi jie fang shi ce liang xie ya ,zui chang yong de liang chong fang fa fen bie shi ke shi yin fa he mai bo zhen dang fa [1]。mai rui PM9000jian hu yi de zi dong hua mo chuang xing ce ya (automated non-invasive blood pressure,NIBP)mo kuai shi yong de shi mai bo zhen dang fa ,ji yuan li shi tong guo qi beng gei xiu dai chong qi ,zhi dao xiu dai ya li zu duan shang bei dong mai xie liu ,ran hou an zhao yi ding de fu du fang qi ,zu duan de dong mai xie liu sui zhao ya li de jian shao he mai bo tiao dong ,zai xiu dai zhong chan sheng mai dong ,xiu dai lian jie dao de ya li chuan gan qi ji lu ji xin hao bing fang da ,zai zhuai huan wei shu zi xin hao ,tong guo chan pian ji ji lu he suan fa chu li de chu shou su ya he shu zhang ya