陈彦:迈瑞PM9000多参监护仪的NIBP模块维修三例论文

本文主要研究内容

作者陈彦(2019)在《迈瑞PM9000多参监护仪的NIBP模块维修三例》一文中研究指出:无创血压是利用非直接方式测量血压,最常用的两种方法分别是柯氏音法和脉搏震荡法[1]。迈瑞PM9000监护仪的自动化无创性测压(automated non-invasive blood pressure,NIBP)模块使用的是脉搏震荡法,其原理是通过气泵给袖带充气,直到袖带压力阻断上臂动脉血流,然后按照一定的幅度放气,阻断的动脉血流随着压力的减少和脉搏跳动,在袖带中产生脉动,袖带连接到的压力传感器记录其信号并放大,再转换为数字信号,通过单片机记录和算法处理得出收缩压和舒张压

Abstract

mo chuang xie ya shi li yong fei zhi jie fang shi ce liang xie ya ,zui chang yong de liang chong fang fa fen bie shi ke shi yin fa he mai bo zhen dang fa [1]。mai rui PM9000jian hu yi de zi dong hua mo chuang xing ce ya (automated non-invasive blood pressure,NIBP)mo kuai shi yong de shi mai bo zhen dang fa ,ji yuan li shi tong guo qi beng gei xiu dai chong qi ,zhi dao xiu dai ya li zu duan shang bei dong mai xie liu ,ran hou an zhao yi ding de fu du fang qi ,zu duan de dong mai xie liu sui zhao ya li de jian shao he mai bo tiao dong ,zai xiu dai zhong chan sheng mai dong ,xiu dai lian jie dao de ya li chuan gan qi ji lu ji xin hao bing fang da ,zai zhuai huan wei shu zi xin hao ,tong guo chan pian ji ji lu he suan fa chu li de chu shou su ya he shu zhang ya

论文参考文献

  • [1].迈瑞PM9000监护仪的结构及常见故障分析与检修[J]. 王莹.  科技视界.2011(03)
  • [2].PM9000监护仪维修电源1例[J]. 黄兰明.  医疗卫生装备.2012(01)
  • [3].迈瑞PM9000监护仪使用注意事项[J]. 黄德生,王宏春.  医疗卫生装备.2007(08)
  • [4].IntelliVue MMS X2监护仪打造全新病人转运模式[J].   中国医疗设备.2010(10)
  • [5].医院监护仪的维护与保养[J]. 于广亭.  医疗装备.2006(02)
  • [6].怎样在临床上用好监护仪[J]. 王桂玲.  医疗卫生装备.2005(09)
  • [7].509B型监护仪黑屏检修[J]. 李宁.  医疗卫生装备.2004(03)
  • [8].浅谈监护仪的临床使用[J]. 胡肖聪.  医疗卫生装备.2003(S2)
  • [9].医用监护仪的临床应用与分类[J]. 袁林.  医疗设备信息.2002(05)
  • [10].PM9000监护仪电源故障检修2例[J]. 向逾,种银保,陈渝,吕思敏.  医疗卫生装备.2016(01)
  • 论文详细介绍

    论文作者分别是来自医疗装备的陈彦,发表于刊物医疗装备2019年17期论文,是一篇关于迈瑞论文,多参监护仪论文,模块论文,维修论文,医疗装备2019年17期论文的文章。本文可供学术参考使用,各位学者可以免费参考阅读下载,文章观点不代表本站观点,资料来自医疗装备2019年17期论文网站,若本站收录的文献无意侵犯了您的著作版权,请联系我们删除。